different modes determines whether the volume or the pressure control mode will be
selected.
The SIMV mode is combined with the PSV mode, and forms SIMV+VCV+PSV,
SIMV+PCV+PSV and SIMV+PRVC+PSV modes.
In the SIMV mode, the following parameters need to be set:
-
【
P
supp
】
-
【
Freq
】
-
【
T
I
】
-
【
V
T
or P
insp
】
-
【
FiO
2
】
-
【
PEEP
】
-
【
P
TRIG
or F
TRIG
】
-
【
T
p
】
-
【
ETS
】
-
【
P
rate
】
2.8.6 DPAP mode
DPAP (Dual Positive Airway Pressure). The ventilator forms the high pressure level and
low pressure level according to the pre-set value, and switches by the pre-set frequency
and inspiratory time. Patients can breathe spontaneously at the high pressure and low
pressure level by trigger, and the ventilator provides support ventilation according to the
pre-set pressure.
In the DPAP mode, the following parameters need to be set:
-
【
T
I
】
-
【
Freq
】
-
【
ETS
】
-
【
FiO
2
】
-
【
P
high
】
-
【
P
low
】
-
【
PS
high
】