55
NIF can be set in SPONT mode.
Touch the option frame to select “Start”, and the NIF is performed automatically, and the
measured value is displayed on the measured value screen.
Parameter range: -40~0cmH
2
O.
10) Automatic tube compensation
It is a compensation for a reduction of work of breathing targeted due to the inserted tube.
NIF can be set in all mode, except NPPV mode.
Touch option frame to select the tube diameter and compensation range, and system will
make a compensation automatically.
Intubation diameter setting range: OFF, 5.0-12mm, increment: 0.5mm, default: OFF.
Compensation range: 0-100%, increment:1%, default: 80%.
11) AutoPEEP
It is a pressure differentials of the setting and measured PEEP.
Touch the option frame to select “AutoPEEP”, and the system is tested automatically, and
the value is displayed on the measured value screen.
Parameter range: 0~100cmH2O.
AutoPEEP is available in all modes.
12) CO
2
module zeroing
For mainstream CO
2
module, they need to establish a zeroing every time you have any
doubts about the accuracy of the measured values.
Touch
【
CO
2
module zeroing
】
touch screen icon, the module will begin normal zeroing.
13) Close the setting screen
Touch
【
Close
】
touch icon to exit the setting screen.
4. Version Information
The version information screen provides the operator with displays of software version
information. It’s shown as below.