6
Cautions:
cause damage or security to the system.
Please refer to the assembling instruction described in the user manual to assemble
the system or the optional accessories.
The gases used in the system must conform to the following standards for
concentrations of water vapor and oil:
Air: H
2
O
<
7g/m
3
, Oil
<
0.5mg/m
3
Oxygen: H
2
O
<
20mg/m
3
Pressure value is given in cmH
2
O:
1kPa = 10cmH
2
O
100kPa = 1bar
≈
1atm
≈
1kgf/cm
2
≈
14.5psi
As general rule, always be careful not to touch the pins of the external electric
connector.
The measuring value condition shown in this user manual is Ambient temperature and
pressure, dry gas (ATPD) condition.
Sharp tools should be away from the screen.
Accumulation of excess liquid in the expiration sensor is not allowed (such as in clean
and disinfection period, otherwise it may affect the function of the ventilator.
When lifting or moving the ventilator system or some parts of the system, please
comply with the instruction of the machine and do some preparation of the safety.
Please do not use the soft tube with the characteristic of antistatic electricity or the
electricity conducting.
The ventilator contains no latex element.
Make sure a resuscitator is always available.
Make sure there is at least one battery as backup power supply.
Once the system is connected to the patient, make sure there should always be
someone professional to monitor the system operation.
Do not disassemble the expiratory module during operating the ventilator.
When the ventilator is powered off, please make sure the humidifier is also powered