41
5.2 Start-up
1. Plug AC power cord into the main AC outlet and make sure the power supply works
normally.
2. Turn on the system switch located on the rear of the equipment.
Warnings:
If abnormal alarms occur when the system starts up, do not use the equipment,
contact the authorized serviceman or the after service department of the Company.
Only the specified, protective grounded power supply can be used in this equipment.
Note:
The current screen displays as DEMO, there may exist difference with the actual
interface of the product. Please refer to the actual interface.
When turning off the equipment, the system will save the current ventilation mode
and relative parameters settings, which can be selected by the user as a patient
category when the system is used in the next time.
5.3 Power-up screen
The power-up screen is the first screen displayed on the screen when the ventilator is
turned on.
In Power-up screen, the system performs the functional check. The check items include
power (10V, 5V, 12V, -12V), inspiratory pressure sensor, expiratory pressure sensor,
oxygen sensor and so on. If the failure occurs, the error information is displayed in the
screen and the system stops running. Please contact the Company Technical Support if
need.
The system continues to the next screen until the failure is solved.