81
B.6 CO
2
module specification (Optional)
Parameter
Specification
Measuring mode
Mainstream
Operation temperature
0-40
℃
/ 32-104
℉
Storage and transportation
temperature
-40-75
℃
/ -40-167
℉
Operating humidity
< 40 hPa H2O (non-condensing)
(95 %RH at 30 °C)
Storage and transportation
humidity
5-100 % RH, condensing
Operation atmospheric
pressure
525-1200 hPa (525hPa corresponding to an altitude of 4572 m/15000
feet)
Storage and transportation
pressure
500 to1200 hPa
Total system response time
<
1 second
Accuracy specifications(during
standard conditions
)
Range
Accuracy
0 to 114 mmHg
±
(0.2 mmHg + 2% of reading)
High Continues Pressure
It will give an alarm when the airway pressure is
continuous above
(
PEEP+15
)
cmH
2
O for 15 seconds
Apnea
Range:10
~
60 s, increment :5 s
O
2
supply pressure low
The O
2
supply pressue is less than 0.28 MPa
Air supply pressure low
The O
2
supply pressue is less than 0.28 MPa
AC failure
The alarm time is more than 120s
Low battery capacity
The battery working time is 10 minutes after alarm
Battery exhausted
The battery working time is 5 minutes after alarm
Alarm silence count down
≤100 s
Alarm sound
>65dB