ProteOn XPR36 System | A ProteOn XPR36 Instrument and Peripherals
206
The primary components of the optical detection system are the light source,
the sensor chip (gold-coated glass prism), the imaging optics, and the charge-
coupled device (CCD) camera. The CCD camera collects the light that is
reflected from the chip surface. Images of the interaction spots are projected
onto the CCD sensor. The optical system has the advantage of scanning
without moving parts.
For more information about Surface Plasmon Resonance, see Appendix C.
Reservoir, Overflow Sensor, and Drain
Pump
A small reservoir is located beneath the wash station to collect fluid (flow-
through, washes, and condensate). A sensor in the reservoir detects when the
reservoir is full and turns on the drain pump for a few seconds to drain the
reservoir contents into the collection tank.
Computer and Display
The ProteOn XPR36 system is supplied with a computer controller running
Microsoft Windows XP operating system, a 19" display, keyboard, mouse,
ProteOn Manager software, and Microsoft Office software.
Содержание XPR36
Страница 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Страница 2: ......
Страница 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Страница 4: ......
Страница 10: ...ProteOn XPR36 System viii...
Страница 166: ...ProteOn XPR36 System Analysis 148...
Страница 188: ...ProteOn XPR36 Maintenance 170...
Страница 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Страница 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Страница 250: ...ProteOn XPR36 System E Ordering Information 232...
Страница 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Страница 276: ...ProteOn XPR36 System Glossary 258...
Страница 287: ......