Processing Data
115
Processing EV Correction Data
Use the controls on the Data screen to select and group the analyte data for
processing. See Data Filters and Display Controls on page 103.
On the Data screen a wizard guides you through selecting reference types,
excluding data points, and choosing calibration steps. Note that each sample
name reflects the calibration step number.
To apply excluded volume correction:
1. Select Channel Reference on the Process menu and the type of channel
reference (Interspot, Column, or Row) to apply to the raw sensorgram
data.
2. On the Process menu select Channel Reference > Correct for Excluded
Volume. Calibration plots appear that show the data best fit line, and
the bottom of the screen displays a wizard. The wizard screen shows
the R
2
value for the fit.
Содержание XPR36
Страница 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Страница 2: ......
Страница 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Страница 4: ......
Страница 10: ...ProteOn XPR36 System viii...
Страница 166: ...ProteOn XPR36 System Analysis 148...
Страница 188: ...ProteOn XPR36 Maintenance 170...
Страница 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Страница 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Страница 250: ...ProteOn XPR36 System E Ordering Information 232...
Страница 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Страница 276: ...ProteOn XPR36 System Glossary 258...
Страница 287: ......