Pronk Technologies
2017 07-18
www.pronktech.com
800-541-9802
Page 22
©2015
Flow Tr ax
Technical Specifications
Pressure Performance
Range: Manual: -12.5 to +75 PSI
Electronic Occlusion (IV pump Testing): 0 - 40 PSI
Accuracy: 0.5% of r/- 1 LSD
Resolution: .01 PSI
Units of Measure: mmHg, PSI (G), inH20, cmH2O, kPa, atm, inHg, BAR (A), BAR (G)
Flow and Volume Performance
Range: 0.5 to 999 ml/hr
Accuracy*: 1% of reading ± 1 LSD for 1ml/hr to 999ml/hr, 2% ± 1 LSD below 1ml/hr. 0.85% of reading
for 500ml/hr, 12ml Volume under steady flow conditions.
Requirement for Accuracy:
Precision
Volume
: ≤ 0.01ml
Precision of FlowTrax is equivalent to 0.01 g based on the conversion of volume (ml) to weight (g) by
calculation using standard temperature and density value of water. Precision was established using a rate
of 500ml/hr and a “dispensed” volume of 10ml.
Flow Rate
Required Volume/Time
.5 to 10ml/hr
0.5ml
10.1 to 199ml/hr
3 minutes (see “Flow Rate and Volume Accuracy - Time vs. Volume”
Section in Op. Manual)
200 to 999ml/hr
10ml
Temperature Performance
Range: -28.9 to +115.5°C
Accuracy: 0.1% +/- 0.2 °C
Resolution: 0.1 °C
Digital Timer / Stop Watch
Range: 0 to 12 Hours
Accuracy: 0.5%
General
Power: Powered by two AA (Alkaline or Lithium) cells or 6 VDC Input
Battery life: Alkaline – minimum 4 hours, Lithium – minimum 12 hours
Size (HxWxD): 5.5 in (13.97 cm) x 3.25 in (8.26 cm) x 2 in (5.08 cm)
Environmental: Operating Temperature: 10 to 35 °C (50 to 95 °F)
Storage and Shipping Temperature: -10 to +40 °C (14 to 104 °F) with water drained.
Humidity: 10% to 80% non-condensing, Altitude: 0 to 6500 feet (2000 meters)