Pronk Technologies
2017 07-18
www.pronktech.com
800-541-9802
Page 13
©2015
Preventative and Routine Maintenance Chart:
Segment
Frequency
Description
Glassware
Monthly
(weekly if
testing very
low rates)
Consists of Glass Tube Assembly found by removing side door of
FlowTrax. Be sure to use “Clean Glass” function prior to removing
Glass Tube Assembly. Remove glass and follow cleaning process
described in Routine Maintenance section of this manual.
Filter
Assembly
Quarterly
Remove filter assembly and follow cleaning process described in
Routine Maintenance section of this manual.
Fluid Path
Annual
Consists of the manifold, occlusion chamber, O rings, transducer,
check valve, interconnect hosing, filter assembly, valves and valve
block. Glass Tube Assembly and Filter Assembly called out
separately. Maintenance of the Fluid Path and associated
components of the FlowTrax system is recommended on an annual
basis, performed as part of the systemic calibration test.
Mechanical
Annual
Consists of stepper motor, optics carriage, belt and pulleys.
Maintenance of the entire mechanical parts of the FlowTrax system
is recommended on an annual basis, performed as part of the
systemic calibration test.
System
Annual
Calibration and maintenance of the entire FlowTrax system is
recommended on an annual basis. The Pronk FlowTrax calibration
procedure must be followed by an authorized calibration company
using NIST traceable fixtures defined in the procedure to guarantee
the device meets all specifications.