Press the ↲ key to start calibrating at
[0.02 NTU]
ð
A 30 second countdown beings, after which the DULCO
®
turb C is calibrated
at
[0.02 NTU]
.
14.
Remove the
[0.02 NTU]
calibration cuvette from the DULCO
®
turb C
15.
Insert the cuvette sample with the measuring cuvette into the DULCO
®
turb C.
16.
Close the in-line flow fitting by screwing it in an anticlockwise direction
ð
The DULCO
®
turb C is now calibrated and is reset to
[AUTO]
mode
Incorrect calibration
A0528
Fig. 34: Display [Incorrect calibration]
If the
[Err CAL]
display appears, it indicates that the internal analysis carried out by the
DULCO
®
turb C has revealed that the calibration was executed incorrectly.
To carry out the calibration procedure again, press the
[MODE/EXIT]
key and restart the
calibration.