Fig. 26: Insert sheet for standard solutions
Handling the standard solution
The standard solution 1000 NTU must be lightly shaken before use for calibration.
Standard solution 10 NTU must be transferred to the cuvette from the storage bottle
before it is used for the first time. The filling for the 10 NTU cuvette is then usable for
24 hours.
Do not use the standard solution once the use-by-date has expired.
The cuvettes for the 1000 NTU and 0.02 NTU standard solutions may not be
opened.