7.
Insert the
[10 NTU]
calibration cuvette into the DULCO
®
turb C so that the marker
‘Indexing the standard solution with your cuvette’
8.
Slowly rotate the calibration cuvette 20° from the centre point to the left and then
20° from the centre point to the right
ð
Leave the calibration cuvette in the position with the lowest measured value.
A0524
Fig. 31: Display [Start calibration ]
9.
Press the ↲ key to start calibrating at
[10 NTU]
ð
A 60 second countdown begins, after which the DULCO
®
turb C is calibrated
at
[10 NTU]
.
Operation
64