4. Hose connector. 4.75 mm interior
diameter, 8 mm external diameter.
Hose connection to the sampling site
5. Detail: Hose
6. Detail: Union nut hose connection
7. Detail: Replacement hose connection
8. Shut-off clamp. To shut-off the inlet in
an emergency or for necessary work
4.3 Installation (electrical)
WARNING!
Live parts!
Possible consequence: Fatal or very
serious injuries
–
Measure: Disconnect the power
supply before opening the
housing
–
Disconnect damaged, defective
or manipulated devices from the
power supply
–
The provision of a suitable iso‐
lating device (emergency-off
switch, etc.) is the responsibility
of the plant operator
NOTICE!
Do not route any RS-485 cables in
the same cable duct as the power
supply cable. This may result in major
anomalies.
The signal leads of the
DULCO
®
turb C may not be routed
alongside faulty cabling. Faults could
lead to malfunctions in the
DULCO
®
turb C.