Prerequisite: The DULCO
®
turb C is ready for operation and is set to
[AUTO]
mode.
Fig. 28: Display [Calibration 1000 NTU]
1.
Select
[CAL]
mode by pressing the
[MODE/EXIT]
key once
ð
An arrow appears next to the
[CAL]
entry in the LCD display (see Fig. 28). The
lower display value (1000) is the selected calibration point. The upper dis‐
played value is the currently measured NTU value.
2.
Open the flow unit by screwing it in a clockwise direction and remove the cuvette
sample together with the cuvette
3.
Insert the
[1000 NTU]
calibration cuvette into the DULCO
®
turb C so that the
marker is facing towards you, see
‘Indexing the standard solution with your cuv‐
4.
Slowly rotate the calibration cuvette 20° from the centre point to the left and then
20° from the centre point to the right
ð
Leave the calibration cuvette in the position with the lowest measured value.