7. Technical Specifications
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 93
7.2.2
AS-active-probes – Overview Normal
For most application our AS-active-probe of class normal are suited. The
fields typically occurring in technical areas can be measured with this
probes.
class
model
type linearity error
(2)
thick
mm
operation
temp. °C
Normal:
AS-NTP 0,6
T
< 0.5 % ±0.2 mT 0.6 ± 0.1
5 – 100
(4)
AS-NTM
T-Ms
< 0.2 % ±0.2 mT 1.4 ± 0.1
5 – 50
AS-NTM-2
T-Ms
< 0.05% ±0.2 mT
1.4 ± 0.1
5 – 50
AS-NAP
A
< 0.5 % ±0.2 mT
Ø 6.0
5 – 50
AS-NTP-Hot-05
T
< 0.2 % ±0.2 mT 1.5 ± 0.1
-40 – 150
(3)
AS-NTP-Flex
T
< 0.5 % ±0.2 mT
up to 1.5 T
0.6 ± 0.1
5 – 100
(4)
AS-NTP-Flex
0,6
T
< 0.5 % ±0.2 mT 0.6 ± 0.1
5 – 150
(4)
AS-NCu-Wire
T
< 0.5 % ±0,2 mT
up to 1.5 T
0.6 ± 0.1
5 – 100
(5)
Table 6
(2) at +20 °C or +25 °C
(3) probe, handle and cable = -40 °C to +150 °C;
probe plug = +5 °C to +50 °C
(4) at first 70 mm = +5 °C to +100 °C;
handle, cable and probe plug = +5 °C to +50 °C
(5) at first 150 mm = +5 °C to +100 °C;
handle, cable and probe plug = +5 °C to +50 °C
(7) at first 70 mm = +5 °C to +150 °C;
handle, cable and probe plug = +5 °C to +50 °C