6. Operation
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 60
6.4.1.2 Axial Probe
A positive reading is obtained when the lines of magnetic field left the black
end face of the axial probe at right angle.
Figure 38 Measurements using an axial probe
The maximum field value is measured when the lines of magnetic field
extend in parallel with the measuring probe. The perpendicular part of the
flux density is not displayed.
Figure 39 Trigonometric function of the axial probe