6. Operation
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 74
At this output the amplified measurement signal is available. At maximum
amplitude the maximal output current is 2 mA. Correspondingly the load has
to be at least 5 k
Ω
. For smaller amplitudes the output can deliver even
higher currents.
The outcome of the given current drive capability is that higher capacitive
loads results in a reduction of bandwidth.
Unloaded the amplifier has a bandwidth of >100 kHz whereby even fast
transient signals can be transferred.
6.5.2.6 Gain Switch
With this switch the gain can be switched between x5 (±2 V
→
±10 V) and
x50 (±0.2 V
→
±10 V).
6.5.3
Adapter Cable
With the AS-probe adapter an adapter cable is delivered. This cable has a
15pol SubD female connector at one site and four single wires at the other
side. Therefore with this cable a AS-active-probe can easily be connected to
the AS-probe adapter.
The cable can be extended at the side of the SubD connector with probe
extension cords as well as at the side of the single wires.
Figure 60 Adapter cable
The assignment of the single wires to the corresponding pins of the probe and
the connectors of the AS-probe adapter can be found in the table below.