6. Operation
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 48
6.3.3.23 Command “version”
command:
version
without
parameter
shows the version of the installed firmware
example:
version
firmware version is V1.0, built Sep 8 2011, 13:10:34
6.3.3.24 Command “zero”
command:
zero [{offset|set}]
without
parameter
shows the current value of the offset compensation
with parameter
With “set” as parameter the automatic offset
compensation process is started.
If a number is given as parameter, it is taken as new
value for the offset compensation. The exact transfer
factor is device dependent.
parameter
set, -39320
≤
offset
≤
26213
see also section 6.3.1.7 Key “zero” – Offset Compensation
examples:
zero
zero compensation value is -7365
zero set
zeroing.....
zero compensation value is 10610
zero -12345
zero compensation value is -12345