7. Technical Specifications
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 112
Figure 90 AS-Adapter 3
Supply
Supply voltage
9 V – 36 V DC
Power consumption
max. 3 W
Output voltage probe supply
±3 V
Output current probe supply
max. 60 mA (20mA each probe)
Signal
Gain
switchable x5,x50
Offset at output X, Y, Z:
in x5: max ±0.25 mV (at 23°C)
in x50: max ±2.5 mV (at 23°C)
Zero drift at output
in x5: max ±0.025 mV/K
in x50: max ±0.25 mV/K
Gain error
typ. ±0.1 %, max. ±0.4 % (DC, at
23°C)
typ. ±0.005 %/°C
Input voltage range
in x5:
±2 V
in x50
±0.2 V
Input resistance
22 k
Ω
Output voltage
max. ±10 V, at load min. 5 k
Ω
Short-circiut proof
Output resistance
<1
Ω
Load resistance
min. 5 k
Ω
to keep specification
Bandwidth (-3 dB)
100 kHz; depends also on used
probe
Isolation
Galvanic isolation
supply – signal: 100VDC, 70VAC