To ease the connection of the AS-active-probe to existing analog inputs with
±10 V input range, the AS-probe adapter contains an integrated amplifier.
This amplifies the output signal of the AS-active-probes from ±2 V to ±10 V.
With a switch, an additionally 10times higher gain can be chosen which
allows to perform even sensitive measurements.
Table 5 shows the measurement ranges as well as the transfer factors for
the analog output resulting from the different probes.
class
ranges and transfer factors
with AS-probe adapter
range x5, x50
High:
x5
x50
(1)
20 T
2 T
10 V / 20 T
10 V / 2 T
Normal:
x5
x50
2000 mT
200 mT
10 V / 2000 mT
10 V / 200 mT
Low:
x5
x50
200 mT
20 mT
10 V / 200 mT
10 V / 20 mT
Ultralow:
x5
x50
200 µT
20 µT
10 V / 200 µT
10 V / 20 µT
Table 5
(1) calibrated up to 12 T
Units
•
T – Tesla
•
G – Gauss
•
Oe – Oersted
•
A/m - Ampere per Meter
For conversion of magnetic units see our application note “PE005 –