7. Technical Specifications
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 101
7.2.11 Axial Probe 2000 mT (AS-NAP)
Figure 80 Size axial probe 2000 mT (AS-NAP)
Ranges (with FM 302)
±20 mT; ±200 mT; ±2000 mT
Effective area
0.2 mm²
Diameter of support
Ø 6.0 mm
Length of support
180 mm
Transfer factor
1 V/T
Bandwidth (- 3 dB)
0 - 35 kHz
Rise time
<3 µs
Linearity error
<0.5 % ±0.2 mT (at 20 °C ±1 °C)
Temperature coefficient
max. -0.05 %/K, typ. -0.03 %/K (0 to 50 °C)
Zero drift
max. ±0.020 mT/K, typ. ±0.010 mT/K (DC)
Noise
typ. 21 µT
RMS
(10 Hz – 10 kHz)
typ. 18 µT
PP
(DC – 10 Hz, 50 s)
Operation temperature
+5 °C to +50 °C
Storage temperature
-10 °C to +60 °C
Max. relative humidity
70 % at +35 °C
Power
±3 V through FM 302, AS-probe adapter,
AS-Adapter 3 or PLC
Connector
15 pol. SubD
Output impedance
<1
Ω
Length of cable
1.5 m
Technical specifications are subject to change without prior notice!