Seite 38
Before lubricating the guide rods of the Proceq ZAA 2600, always switch off
the apparatus and unplug it.
No oil shall be sprayed into the apparatus. The function of electrical or
mechanical components could be impaired.
For lubrication proceed as follows:
•
Remove the inserted plate (see also chapter 8
on page 35).
•
Remove both side panels (17) with the Allen key.
•
Turn the apparatus carefully on one side so that the side opening of the cover
becomes accessible.
•
Put some drops of sewing machine oil on your finger and spread it over the
whole length of the guide rods.
•
Repeat this for the other side of the apparatus.
•
Adjust the start stop (1) at the very beginning and the end stop (2) at the very
end and let the apparatus run back and forth several times, so that the oil
gets evenly distributed over the whole length of the rods.
•
Remount the side panels (17), the one with slit on the right side, the other
one on the left side.
•
Place the plate again.