:SYSTem:TIME <DNPD>,<DNPD>,<DNPD>
:SYSTem:TIME?
Description:
This commands allows set system device time (RTC).
Parameters:
<DNPD>
Hours, Range 0 ... 23
<DNPD>
Minutes, Range 0 ... 59
<DNPD>
Seconds, Range 0 ... 59
Remarks:
Overlapped command
Example:
SYST:TIME 10,45,15
SYST:TIME? Response: 10,45,15
:SYSTem:VERSion?
Syntax:
:SYSTem:VERSion?
Description:
This query retrieves version of implemented SCPI language
Parameters:
<CPD>
SCPI language version
Remarks:
Overlapped command
Example:
SYST:VERS? Response: 1999.0
:UNIT:TEMPerature
Syntax:
:UNIT:TEMPerature <CPD>
:UNIT:TEMPerature?
Description:
This function allows to set unit for all temperature functions (Platinum, Nickel).
Parameters:
<CPD>
{CEL|FAR|K} (default CEL)
·
CEL
degrees of Celsius
·
FAR
degrees of Fahrenheit
·
K
Kelvin
Remarks:
Overlapped command
Value is set according to “Startup” preset on restart and to default on *RST command
Example:
UNIT:TEMP CEL
UNIT:TEMP? Response: CEL