6.3.
LAN Interface (option)
LAN Interface allows communication with decade box using Telnet protocol. A proper setting must be
established.
Following equipment is required:
!
MC642 decade box with LAN, USB, IEEE488 bus option
!
Personal Computer (or other controlling device) with LAN interface
!
LAN cable
The LAN interface must be selected from decade system menu to be in operation (
SETUP->Interface-
>Active bus
). There are following LAN settings accessible from decade system menu under
SETUP-
>Interface->LAN Settings
path (values are default ones):
DHCP
ON
IP Address
192.168.001.100
only valid if DHCP is OFF
Subnet mask255.255.255.000
only valid if DHCP is OFF
Default gateway
255.255.255.255
only valid if DHCP is OFF
Port number
23
Host name
MC642_SN620031
only valid if DHCP is ON
If DHCP (Dynamic Host Configuration Protocol) is enabled, the IP Address and all necessary settings are
done automatically and connection in Telnet protocol is done via “Host name” and “Port number”.
Otherwise the IP address, Subnet mask and Default gateway should be properly set. In this case connection
is done via “IP Address” and “Port number”.
Connection to decade box using Microsoft Telnet terminal with DHCP option enabled:
Figure 26 LAN connection 1
If connection is successful following screen will appear:
Figure 27 LAN connection 2