RS232 communication baud rate setting. The same baud rate must be used in the controller.
GPIB Address
GPIB address setting. Each instrument connected to the GPIB bus must have a unique address.
LAN Settings
Ethernet parameters setting. Device use Telnet protocol. Default setting is:
DHCP
ON
IP Address
192.168.001.100
only valid if DHCP is OFF
Subnet mask255.255.255.000
only valid if DHCP is OFF
Default gateway
255.255.255.255
only valid if DHCP is OFF
Port number
23
Host name
MC642_SN620031
only valid if DHCP is ON
4.8.
Calibration mode
In this mode resistance elements of the decade can be recalibrated. Access to the calibration mode is
from the setup Menu.
Correct password must be entered before calibration. Without correct password the access to the
calibration mode is refused. Default factory set calibration code is “2”. Return to standard mode is possible
after pushing the key ESC.
Figure 22 Password entry
Recalibration procedure consists of measuring of 43 basic resistance values and entering their actually
measured data. Calibration point can be changes using display soft keys “Previous” and “Next”. Calibration
value of selected resistance can be change using cursor keys
▲
,
▼
,
◄
,
►
.
Following table describes nominal values of calibration points and requested recalibration accuracy: