*OPC
Syntax:
*OPC
Description:
This command sets the OPC bit in the ESR (Event Status Register) when all pending operations
are complete.
Parameters:
None
Remarks:
Overlapped command
Example:
*OPC
*OPC?
Syntax:
*OPC?
Description:
This command returns “1” to the output queue after all pending operations inside decade box
are complete.
Parameters:
<DNPD>
always returns 1
Remarks:
Sequential command
Example:
*OPC? Response: 1
*OPT?
Syntax:
*OPT?
Description:
This command return the instrument’s hardware fitment. The only parameter returns presence
of GPIB/LAN/USB interface.
Parameters:
<DNPD>
0 – extended interface not present, 1 – extended interface present
Remarks:
Overlapped command
Example:
*OPT? Response: 1
*WAI
Syntax:
*WAI
Description:
Prevents the instrument from executing any further commands or queries until all previous
remote commands have been executed.
Parameters:
None
Remarks:
Sequential command
Example:
*WAI