5. Performance verification test
Parameter verification procedure is described in the chapter. Verification procedure is based on
measuring resistance on the decade output terminals with standard multimeter in recommended points.
Required equipment
!
Ohm-meter nominal accuracy 0.001% in range 100 m
to 20 M
(type Fluke 8508A or similar)
Decade setting
Switch decade to the resistance function. Connect standard multimeter to the decade output terminals. Use
four-wire connection technique.
Procedure
Use following procedure to perform parameter verification test:
1.
Switch both instruments on and let them for 1 hour stabilize in the laboratory with ambient temperature
23
!
3 °C. Connect resistance decade terminals R4W to the standard ohm-meter (multimeter).
2. Case of decade should be grounded or connected to the Lo terminal of multimeter.
3. Check resistance value in points according to Table I.
Maximal absolute deviations MC642
Nominal value
MC642
max. deviation
180 m
15 m
300 m
15 m
700 m
15 m
1.3
16 m
2.5
16 m
5.0
18 m
9.5
20 m
19
25 m
36
33 m
70
50 m
140
85 m
250
50 m
500
100 m
1 k
200 m
2 k
400 m
4 k
800 m
8 k
1.6
16 k
3.2
40 k
8
80 k
16
150 k
30
300 k
60
700 k
140
1.5 M
300
3 M
1.5 k
6 M
3 k
Table 2 Verification - allowed deviations
26
Operation manual