31
MA-Di
COMPRESSED AIR AT ITS BEST
8.2.2.3
Configure program-controlled counter reset
rTo
Select, continue with
a)
or
b)
a)
Manually reset counter
Set
Press until
rES.T
is displayed, then briefly press
Mode / Enter
.
b)
Enter value for time-controlled reset
Set
Press until the desired value is displayed (intervals from 1 hour to
8 weeks), then briefly press
Mode / Enter
Set
Press until
rES.T
is displayed, then briefly press
Mode / Enter
.
8.2.2.4
Deactivate counter reset
rTo
OFF
Select and
set.
The counter is only reset after overflow (= factory setting).
Overflow: After the maximum value (9 999 999 Nm³), the counter is
reset to 0.
8.2.2.5
Configure counter reset by external signal
OU2
InD
Select and
set.
Din2
Select and set reset signal.
HIGH
= Reset with high signal
LOW
= Reset with low signal
+EDG
= Reset with rising flank
-EDG
= Reset with falling flank
The LED 7 (
6.1. Operating and display elements | p. 22) shows the input status also
in the case of an active external reset.