- the preset value is to be checked or a new value of the fixed  threshold is to be set for the searching 

mode "0-9"  (the  threshold  setting  range  constitutes  1.0  –  99.0  s

-1

,  recommended  –  1.5 s

-1

)  (

РМ

1703GN

О

РМ

1703GN

А

РМ

1703GN

В

); 

-the preset states of indicators of sound and/or vibrating ones are to be set, or thay are to be replaced 

(on/off); 

 

-the preset sound volume of the sound indicator is to be checked or changed. 

 

 

GAMMA-CHANNEL N-FACTOR SETTING

 

For setting  

n-factor

 one should be shifted to the mode of settings, for what  a MODE push button 

should be pressed and kept pressed for more than 5 s. "CAL." will appear on the LCD, and then the installed 
value of n-factor of the gamma channel. 

 

 

 

For changing the  value of n-factor one should shortly press LIGHT push button during subsequent 

four seconds. The preset value of n-factor will blink by indicating  its possible change. If the LIGHT button 
was  not  pressed  during  the  said  time  interval,  the  indicator  automatically  returns  to  the  working  mode. 
Subsequent  pressings  of  the  LIGHT  push  button  increase  the  preset  value  of  n  factor    with  pitch  0,1. 
Subsequent pressings of the MODE push button increase the preset value of n factor  with pitch 0,1. If the 
push  buttons  are  kept  pressed  the  values  are  reduced  or  increased  accelerated  with  the  same  pitch.  After 
setting the required n-factor of the gamma channel after expiry of approximately 6 s after the last pressing 
the push button the indicator will automatically be shifted to the calibration mode.  

 
 

SETTING OF 

N FACTOR OF NEUTRON CHANNEL

2

1

 

For setting 

n-factor 

one should shift to the mode of settings, for what  a MODE push button should 

be pressed and kept pressed for more than 5 s.  "CAL." will appear on the LCD, and then the installed value 
of n-factor of the gamma channel.  A MODE push button should one be pressed and the value of n-factor of 
the neutral channel will be indicated on the LCD. 

For changing the  value of n-factor one should shortly press LIGHT push button during subsequent 

four seconds. The preset value of n-factor will blink by indicating  its possible change. If the LIGHT button 
was not pressed during the said time interval, the indicator automatically returns to the working mode.  

                                                           

2

 

РМ

1703GNO, 

РМ

1703GNA, 

РМ

1703GNB

 

Содержание PM 1703GNA

Страница 1: ...POLIMASTER PERSONAL RADIATION DETECTORS 1703 1703GN 1703 1703GN 1703 1703GN 1703 1 1703 1 OPERATION MANUAL http www polimaster com...

Страница 2: ...Indication mode of neutrons 17 Indication mode of the average speed of and count of neutrons over the time of accumulation 17 Quick switching of sound or vibration indication 17 Communication mode by...

Страница 3: ...erred to as MED The indicator is not a meter The indicator can be operated both in premises and in the open air The indicator can be used by a wide circle of consumers who are by their mode of activit...

Страница 4: ...oth type Possibility of identification of the radio nuclide composition of the substance with the aid of ATTENTION The indicators may differ by the totality of included activated operation modes Opera...

Страница 5: ...1 Personal radiation detector 1703GNO 1 Personal radiation detector 1703GNA 1 Personal radiation detector 1703GN 1 Disk software on CD 1 1 1 1 1 1 1 1 Panasonic POWER LINE LR6 power supply cell 1 1 1...

Страница 6: ...neutron radiation registration 1703GN 1703GN 1703GN From heat to 14 0 MeV Range of indication of MED of photon radiation 1703 1 1703 1A 1703 1703GN 1703 1703 1703GN 1703GN 0 01 9999 mc v h 1 1000 mR h...

Страница 7: ...corded events to the indicator storage by 1000 Terms of operation range of ambient temperatures relative humidity From minus 30 C to 50 22 F to 122 F LCD from minus 15 C to plus 50 by 95 at 35 C 95 F...

Страница 8: ...quirements of IEC 60079 0 owing to the following designing and circuit technical solutions provision of the normal degree of mechanical strength of the indicator casing in accordance with the requirem...

Страница 9: ...irected inside the indicator the cover of the supply section is to be restored When the supply cell is installed the indicator is automatically switched on During switching on and during operation of...

Страница 10: ...ion test CAL OL OFF 1 3 etc an indication type sound or vibration 6 the analogue scale consisting of 19 segments is intended for time indication to completion of the internal processor tests reduction...

Страница 11: ...t the LCD for several seconds After completion of the tests the indicator must be shifted to the calibration mode by the background level except for devices having the search mode 0 9 the analogue sca...

Страница 12: ...ing analogue scale Tests are performed The time to completion of the test is displayed in relative units on the analogue scale in the form of the reducing number of the indicated segments After the te...

Страница 13: ...function may be switched on in the mode of communication with the PC Automatic calibration makes it possible to automatically retain the high sensitivity at reduction of the background level and avoid...

Страница 14: ...he threshold value indication light sound and or vibrating is switched on 3 The operation threshold for the neutron channel21 c The minimum detection level calculated in the calibration mode and takin...

Страница 15: ...ization can be continued If necessary these actions cam be repeated several times up to location of II During localization of the source of neutron2 or mixed gamma and neutron radiations the sound and...

Страница 16: ...ation source 2 The safety threshold the 2nd threshold by MED is only set in the mode of communication with a PC by the user administrator The range of threshold setting is equal to 10 7000 R h 0 1 70...

Страница 17: ...E SPEED OF NEUTRON RADIATION INDICATION DURING ACCUMULATION 2 In this mode in the upper line of LCD the average count time is shown during accumulation time s 1 The lower LCD line shows the value of t...

Страница 18: ...the indicator enters after approximately 1 minute the sleeping mode and being so LCD will have the following forms the radio channel is on no communication with the P the radio channel is on communica...

Страница 19: ...LIGHT push button increase the preset value of n factor with pitch 0 1 Subsequent pressings of the MODE push button increase the preset value of n factor with pitch 0 1 If the push buttons are kept pr...

Страница 20: ...of the searching threshold by MED will blink by indicating its possible change If the LIGHT button was not pressed during the said time interval the indicator automatically returns to the working mode...

Страница 21: ...ode MODE OF INDICATION SETTINGS ON OFF OF SOUND AND OR VIBRATING INDICATION The state selection on off of vibrating and sound indicators from the front panel is possible if this mode is allowed at set...

Страница 22: ...ce then at entering the mode of sound volume setting Aud note will be indicated on the LCD as well as the preset sound indicator value from 1 to 5 For changing the sound volume a LIGHT push button sho...

Страница 23: ...omatic start up is supported Operation with the PP is described in th Help site in the attached textual document which are installed along with the user s software For switching on the communication m...

Страница 24: ...tion is performed by wiping with clothes moistened in ethyl spirit replacement of a supply cell in the explosion hazard area is strictly prohibited In case of visible mechanical damages of the casing...

Страница 25: ...anufacturer s permit The warranty does not concern the indicators at expiry of the warranty operation term if the indicator is purchased within the limits of the warranty storage period in case of the...

Страница 26: ...nconformity for which Polimaster is responsible Purchaser shall inform Polimaster of the nature of such defect or nonconformity in reasonable detail and shall request authorization from Polimaster to...

Страница 27: ...to modes 1 2 or 3 5 Mode 0 9 Mode of relative indication 0 9 Can be ON autonomous or combined with modes 2 or 3 6 Auto calibration Impossible in mode 5 7 Search in measurement mode Possible only in co...

Страница 28: ...t threshold count neutrons Only for mode 5 threshold 1 0 99 0 s 1 with interval 0 5 advisable 1 5 s 1 Additional settings 14 Interval of history record min 15 Resolution of coefficient variation n 16...

Отзывы: