functioning properly. In case of return of the apparatus, we
will only accept original packaging.
5.2
INITIAL USE
Before filling the bath make sure that the bath is clean.
Place the bath spirit level, and connect it to the mains
after the mains-switch is placed in “off” position. Use a
mains-supply with proper grounding.
Important: Never use the bath without fluid
Following applies cooling equipment (TLC and TLB):
When the apparatus has been tilted or transported,
allow the bath a few hours for stabilization.
Do not switch on power, because lubrication-oil inside the
compressor system has to run into the capillary.
It will take several hours before the oil has flown back into
the compressor unit. When the apparatus is immediately
turned on after it has been tilted,
damage may occur to the compressor unit due to
insufficient lubrication.
5.3
PLACEMENT AND VENTILATION
The bath has to be placed in a well ventilated area.
Do not place the bath in a dusty environment. Dust will
block the condenser unit inside the apparatus which will
lead to severe mechanical damage.
Air must circulate freely. Keep 10 cm of free space at both
sides and 15 to 20 cm at the rear of the bath.