8.3.2 FRONT PANEL KEYS
The front panel layout shows 3 operating keys:
The “Page” Key offers following:
-
Temperature readout in °C,
-
Temperature set point in °C,
-
Changing the tuning (PID) parameters,
The “Up and Down” keys allow changing the listed value.
All changed values like set point and PID parameters will
be kept in memory even after switching of the power
supply.
The front panel layout shows the following 3 operating
keys:
-
Set point
-
Offset**(press: <-5.00 .. +5.00°C step 0.01°C)
-
Max Power (press: low, med, hi, max)
-
Boost heater (press on / off)
-
Time const (press: fast, medium slow, precise)
-
Stirrer *
-
Low alarm *
-
High alarm *
-
PID parameter:
(PID set 1, PID set 2. PID set 3, PID set 4)
Proportional band (Pb=1/P where P is
proportional value)
Integral value
Differential value
-
Backlight of display
-
Temp units
-
Baudrate
*
optional
**
This value is added to the measured
temperature by the microcontroller. This way the
temperature readout can be synchronized with
an external calibrated measuring device like the
TT3.
note:
Display switches back automatically to
temperature readout. Changed values are
stored. Or Press to confirm when settings are
altered.
Each time is pressed another option is
displayed.
Up
and
Down
keys allow changing the listed value. All
changed values, like set point and PID parameters, will
be kept in memory when pressing the page key to leave
Page
Down
Up