10.4 DRAINING BATH FLUID
Before removing flammable bath liquids take the
appropriate fire hazard precautions against these liquids.
When draining the bath fluid, be sure that the bath fluid
has cooled down to ambient temperature. If viscosity is
too high at ambient temperature to remove the fluid,
make sure that the hose to drain and waste container
can withstand the temperature of the bath fluid.
When removing bath fluid do not inhale toxic vapor.
Always use appropriate ventilation.
10.5 USING THE DRAIN TAP
The TC baths can be drained via the drain tap located at
the backside of the apparatus.
The drain has an inner diameter of 3/8” gas fitting. The
drain can be connected onto a pipe, tube or drainage
system.
For safety reasons the tap can only be opened by using
a screwdriver. The thread inside the tap is 3/8".
By screwing the valve with a screwdriver 90 degrees anti
clock-wise the valve is fully open and the bath can be
emptied.
USE SCREWDRIVER
TO OPEN TAP
3/8" BSP inside thread