PID CONFIGURATION
The two graphs below show the influence of the integral
setting. The graph on top has a setting of 200 for the
integral parameter. The bottom graph has a setting of 16.
The value of 200 causes less stable control.
I = 200
I = 16
T
em
p
er
at
u
re
[°
C
]
Max
0:00
0:30
1:00
Delta = 0,075°C
± 0,038°C
Min
T
em
p
er
at
u
re
[°
C
]
Max
0:00
0:30
1:00
Delta = 0,048°C
± 0,024°C