7.4
EXTERNAL PUMP
To ensure uniform temperature control all models have a
built in stirring mechanism. Additional to the stirring, the
TLC and TTU (can) offer the possibilty to circulate fluid
trough an external closed system
Both stirrer and pump are protected bij a fuse. This motor
protection fuse is located on the backpanal of the motor
housing
When pumping viscous liquids the thermal fuse can be
activated. The motor fuse can be reset by pressing the
button.
The inlet and outlet of the pump are fitted with 8 mm hose
connections and 1/4” threaded fittings for metal tubing.
The outlet is closed by a small disc, stopping the
circulation action.
Before the fluid can begin circulating through external
systems this disc must be removed by unscrewing the
fitting and hose connection removing the disk.
REMOVE DISK
WHEN
CIRCULATING
PRESS
BUTTON TO
RESET MOTOR
FUSE