If the power of sterilization system is abnormally shut off during the sterilization process, the following
screen will appear. To ensure user safety, the purification process will be active it can take from a few
minutes to tens of minutes, after the status will be ready. Even the purification is well done, you can find
the residual sterilant.
WARNING! POSSIBLE NON-STERILE DEVICE.
Any error may result in a non-sterile medical device, and it requires to restart
a new sterilization cycle with rewrapping using new packaging material.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Any error may result in failure of removing hydrogen peroxide. Wear latex,
PVC (vinyl), or nitrile gloves when handling the loads and cassette.