Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
28
5.3. Preparing the load
Confirming that sterilizing conditions were achieved during a sterilization cycle is one of
important part of the sterilization process. Biological indicators are one way to ensure that
your sterilizer is operating correctly.
Place a biological indicator in the STERPACK
®
, STERPACK
®
Plus or STERLINK
®
chamber,
respectively for the POUCH, POUCH Plus or CHAMBER mode. This biological testing is
recommended to be performed at least once per day or specified by your facility’s policy.
While the sterilizer is warming up, you can use this time to prepare the load. Refer to the
chapter detailing the load preparation information.
Mode
Item description
POUCH
Total mass of the items should be less than 0.5 kg.
POUCH Plus
Total mass of the items should be less than 1.5 kg.
CHAMBER
Total mass of the items should be less than 5 kg.
WARNING! POSSIBLE NON-STERILE DEVICE.
Improper sealing of the STERPACK
®
, STERPACK
®
Plus for the POUCH, POUCH
Plus mode may result in either a non-sterile device or cycle cancellation. Seal tight the
STERPACK
®
by using a recommended sealing machine.
CAUTION! DO NOT REUSE.
The products (STERPACK
®
, STERPACK
®
Plus, STERLOAD
®
) are intended for single
use only. Do not reuse, if you reuse the product an error will occur.
CAUTION! CHECK EXPIRY DATE.
When you use the sterilant cassettes (STERPACK
®
, STERPACK
®
Plus, STERLOAD
®
),
check the expiry date. If the date is expired, dispose and use new available sterilant
cassettes. The use of expired date cassette will occur an error.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle
cancellation or error occurrence. Hydrogen peroxide liquid may be present on the load
or in the chamber.