Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
53
9.4. Sterilant cassette errors
The STERLINK
®
FPS-15s Plus sterilization system is designed to initiate automatically the
sterilization cycle according to the barcode information provided by the sterilant cassettes. The
predictable error source is presented with the intuitive figures, and the user should identify the
source and confirm. If the problems persist, contact local Plasmapp customer service
representative.
Figure 9.4 GUIs for the errors related with the sterilant cassettes
Number
Name
Description
7
Confirm
The user should confirm the reason of the error and press Confirm
button.
8
USED CASSETTE
Probable cause: The used cassette is tried to be reused.
Suggestion: Try again with a new cassette.
9
EXPIRED
CASSETTE
Probable cause: The expired cassette is tried to be used.
Suggestion: Try again with a new cassette or check time setting.
10
INVALID
CASSETTE
Probable cause: The barcode of the cassette is invalid.
Suggestion: Try again with a new cassette.
11
BARCODE NOT
DETECTED
Probable cause: The cassette is not detected.
Suggestion: Clean the barcode window or try again.