![Plasmapp Sterlink FPS-15s Plus Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/plasmapp/sterlink-fps-15s-plus/sterlink-fps-15s-plus_user-manual_1565811031.webp)
Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
31
Notes.
∎
If the cassette is reused or expired one, it results in cycle cancellation.
∎
Improper loading of the cassette may result in scanning error, and the cycle will be
canceled.
∎
When the chamber door is not properly closed, it results in cycle cancellation.
∎
Cycle stop may not obtain sterility assurance level and users can be exposed to residual
sterilant which can be left in the cassette. Ensure protective gloves are worn when
handling an used cassette.
WARNING! POSSIBLE NON-STERILE DEVICE.
Improper loading of the cassette may result in either a non-sterile device or cycle
cancellation. The cassette should be precisely loaded on the cassette loading block of
the sterilizer and properly locked.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle
cancellation or error occurrence. Hydrogen peroxide liquid may be present on the load
or in the chamber.