![Plasmapp Sterlink FPS-15s Plus Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/plasmapp/sterlink-fps-15s-plus/sterlink-fps-15s-plus_user-manual_1565811047.webp)
Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
47
The sterilizer interior chamber can be cleaned with a soft cloth or damp soft cloth with
no running water. When cleaning the sterilizer interior chamber, follow the guidelines.
1.Turn off the power of the sterilizer before cleaning the interior chamber.
2. Wait until the interior chamber has cooled down.
3. Never allow cleaning solution or water to enter the interior or chamber.
Use only dry or damp cloth to gently clean the surface of the interior chamber.
4. Wait until the interior chamber is completely dry before staring the sterilizer.
Notes.
∎
When a user attempt to clean the chamber, door, or interior surface, the user must wear
gloves and follow the guidelines.
∎
Failure to follow the guidelines may result in damage to the sterilizer and may void the
warranty.
WARNING! STERILIZATION SURFACES.
At the end of a cycle, the interior of the sterilizer may be hot. Do not touch the inside of
the chamber or door with your bare or gloved hands. Allow the sterilizer to cool before
touching interior surfaces.
WARNING! HAND CRUSH HAZARD.
The door of the sterilizer is designed to be manually opened and closed, and your hand
can be injured by the door. Keep hands clear.
CAUTION! POSSIBLE DAMAGE ON THE STERILIZER.
Do not use the steel brush (wire brush), steel foam (steel wool), compound and
detergent containing chloride compound. They may damage the sterilizer or its
electrodes. Please remove all detergent, fabric or thread left in the chamber after
cleaning.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle
cancellation or error occurrence. Hydrogen peroxide liquid may be present on the load
or in the chamber.