Rev. 03 / Jan. 2019
Plasmapp Co., Ltd.
STERLINK
®
FPS-15s Plus User Manual
54
9.5. Other system errors
Many system errors are incorporated with the problems with the loads, but the STERLINK
®
FPS-
15s Plus sterilization system also can have a problem resulting from a component failure. If the
problems persist, contact Plasmapp or local Plasmapp customer service representative.
Figure 9.6 GUIs for the system errors
Number
Name
Description
12
Confirm
The user can confirm this system error after contacting customer
service representative.
13
SYSTEM ERROR
Probable cause: A component failure results in preheating error.
Suggestion: Contact the customer service or restart the system.
14
SYSTEM
STABILIZING
Probable cause: A component failure results in completing the cycle.
Suggestion: Wait until recovering system and try again.
Note.
∎
Repairs and adjustments should only be made by the trained and authorized personnel.
∎
When the door cannot be opened after sterilization cycle, please refer to the Section 6.3
featuring “Additional utility menu” for the user to temporarily open the door. If the
problems persist, contact your local Plasmapp customer support representative.