Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
HE1092
The unit has created a new user profile
automatically.
The unit has created a new user profile
automatically. Check your user settings.
HE1093
Failed to identify patient.
The unit failed to identify the RFID of the
patient.
HE1094
Patient already identified.
The patient RFID was rejected, because
the treatment session of the previous
patient had not been ended yet.
HE1095
Patient identification required.
Show the patient RFID to the PlanID
reader.
HE1096
Patient identification not enabled.
The unit has detected a patient RFID, but
patient identification is not enabled in user
settings.
HE1097
RFID unknown.
The RFID is unknown to the system.
HE1098
SSL required.
SSL required. Enable SSL to pass
confidential data between the unit and
Romexis.
HE1099
Unit settings have been modified.
Unit settings have been modified. Reboot
the unit.
HE1100
Invalid Romexis username.
Romexis username of the user contains
characters not supported by the unit.
HE1101
Adding a user to Romexis not allowed.
Romexis does not allow creating new
users. Check Romexis configuration.
HE1102
Reassigning PlanID not allowed.
Romexis does not allow reassigning the
PlanID tag. Check the user settings in
Romexis.
HE2000
Wrong scaler module installed.
This scaler does not have the correct
module installed. Install the correct scaler
module.
HE2001
Unsupported instrument.
This instrument is not recognised.
Software upgrade may be needed.
HE2002
Unsupported assistant syringe.
This instrument is not recognised.
Connect only supported instruments.
HE2003
Unsupported assistant instrument.
This instrument is not recognised.
Connect only supported instruments.
HE2012
Instrument hose memory checksum error. Invalid checksum read from the
instrument hose memory.
HE2013
Syringe in unsupported slot.
The syringe is connected to a slot that
does not support syringes. Consult the
user's manual.
HE2018
Wrong instrument connected to the slot
with Bien-Air MX2 electronics.
Only the Bien-Air MX2 instrument can be
connected to a slot with Bien-Air MX2
electronics. The slot is marked with a
label. Consult the user's manual.
HE2026
Reverse drive not possible because of
torque drive mode.
Micromotor cannot be driven in reverse
direction because torque drive mode is
either autoreverse or autoforward.
31 Help and error messages
User's manual
Planmeca Sovereign Classic 255
Содержание Sovereign Classic
Страница 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Страница 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Страница 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Страница 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Страница 290: ...Optimal space requirements 34 2 5 Side view 34 Technical information 280 Planmeca Sovereign Classic User s manual...
Страница 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Страница 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Страница 296: ......
Страница 297: ......