3. Flexy-holder: Disinfect the supplementary holders, suction tube holders
(including rollers) and instrument holder in a washer-disinfector at 93°C.
4. Flexy-holder: Wipe the holder for the intraoral scanner with Planmeca
approved surface disinfectant.
30.6.4 Weekly cleaning procedures
30.6.4.1 During working day
About this task
Clean the suction system with Dürr MD 555 cleaner 1 - 2 times a week to
prevent the buildup of deposits in the suction system, especially if you are
using dental air polishers.
This procedure is mandatory for dental units with a Dürr amalgam separator
or Dürr VS/A separator.
For more information on Dürr MD 555 cleaner, please visit
Steps
1. Pour 50 ml of Dürr MD 555 cleaner into a rinsing bottle (for example
OroCup). Add 1 l water and mix well.
2. Place the rinsing bottle on a flat surface (table or floor).
3. Remove the suction handpieces from their holders and push them onto
the inserts inside the cap of the rinsing bottle.
30 Cleaning and disinfection
248 Planmeca Sovereign Classic
User's manual
Содержание Sovereign Classic
Страница 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Страница 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Страница 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Страница 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Страница 290: ...Optimal space requirements 34 2 5 Side view 34 Technical information 280 Planmeca Sovereign Classic User s manual...
Страница 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Страница 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Страница 296: ......
Страница 297: ......