The modes are:
Auto stop
The micromotor stops when the torque limit is
reached.
Auto reverse
When the torque limit is reached, the micromotor will
operate counter-clockwise.
Auto forward
When the torque limit is reached, the micromotor will
operate counter-clockwise (auto reverse) for 2
seconds and then return to clockwise direction.
3. Adjust the micromotor speed rate.
3.a. Press Program.
3.b. Press RPM limit.
A pop-up window opens.
3.c. Adjust the micromotor speed rate (RPM).
Alteration of RPM speed rate in different ranges
Range
In steps of:
100 - 200
10 (e.g. 110, 120 etc...)
200 - 500
20 (e.g. 220, 240, 260 etc...)
500 - 1 000
50 (e.g. 550, 600, 650 etc...)
1 000 - 2 000
100 (e.g. 1 100, 1 200 etc...)
2 000 - 5 000
200 (e.g. 2 200, 2 400 etc....)
5 000 - 10 000
500 (e.g. 5 500, 6 000, 6 500 etc...)
10 000 - 20 000
1 000 (e.g. 11 000, 12 000 etc....)
20 000 - 38 000
2 000 (e.g. 22 000, 24 000 etc....)
3.d. Press OK.
The pop-up window closes and the new value is displayed on the
control panel.
4. Adjust the torque limit rate.
4.a. Press Program.
26 Programming
184 Planmeca Sovereign Classic
User's manual
Содержание Sovereign Classic
Страница 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Страница 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Страница 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Страница 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Страница 290: ...Optimal space requirements 34 2 5 Side view 34 Technical information 280 Planmeca Sovereign Classic User s manual...
Страница 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Страница 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Страница 296: ......
Страница 297: ......