To remove the oil collector from the instrument console, unscrew the oil
collector attachment knob and carefully pull the collector out.
30.5 Cuspidor
30.5.1 Bowl
The glass bowl should be cleaned after every patient and after the working
day.
Empty and clean the bowl filter after each patient and replace it when
necessary.
Clean the bowl as follows:
1. Clean the bowl with a soft brush.
2. Rinse the bowl by pressing the Bowl rinse button on the control panel (or
by using the foot control).
3. The outside of the bowl can be wiped clean or disinfected with a cloth.
30 Cleaning and disinfection
242 Planmeca Sovereign Classic
User's manual