Code
Explanation
Comments
H144
Height movement is not possible
because one (or more) of the
positioning control buttons is stuck.
Release the button / joystick.
H148
Height movement is not possible.
The position of the patient support
table is too high.
Press the X-ray unit down
button to move the patient
support table down.
H149
Height movement was stopped
because the C-arm cannot be
moved lower.
Clear any obstruction before
continuing.
H150
Height movement was stopped
because the patient support table
cannot be moved lower.
Clear any obstruction before
continuing.
H151
Line voltage
The line voltage was too low during
exposure.
Exposure was interrupted.
Contact your service
technician for help.
H152
The line voltage is too low.
Exposure is not possible.
Contact your service
technician for help.
H161
Temperature
The temperature of the tube head is
too high.
Wait for a few minutes for the
tube head to cool down.
H162
The temperature of the lift motor is
too high.
Wait for a few minutes for the
lift motor to cool down.
H165
The temperature of the tube head is
too high for the selected exposure
values.
Wait for a few minutes for the
tube head to cool down.
H166
The maximum tube head energy
level was exceeded.
Wait for a few minutes for the
tube head to cool down or
use lower exposure values.
H170
User related
messages
Wrong licence code.
Check the licence code.
H171
The sensor is not attached properly
to the C-arm.
Attach and / or lock the
sensor in position.
H172
The sensor is not attached properly
to the cephalostat.
Attach and / or lock the
sensor in position.
H175
PC program selection is in conflict
with the selected X-ray unit program.
Select another exposure
mode in Planmeca Romexis.
H180
DEC is not available.
H181
The imaging process was cancelled
in Planmeca Romexis.
H182
Timeout in image data transmission. Exposure was interrupted.
Contact your service
technician for help.
H183
The attached sensor is not suitable
for the selected program.
Change the sensor.
H184
Remove the 3D sensor.
H185
The 3D sensor is not attached
properly.
Attach and / or lock the
sensor in position.
12 Help messages
84 Planmeca ProMax
User's manual