NOTE
The function is available for programs that have been activated in menu
2300 Licences.
•
To permanently adjust preset exposure values:
1. Select Program > 2100 Programs.
2. Select a program group (e.g. 2110 2D Panoramic).
3. Select a program type (e.g. Interproximal).
4. Select the exposure values you wish to adjust (e.g. 70 kV / 10 mA for
patient size M).
•
In 2D panoramic programs select also the MultiView button if you
wish to adjust the presets for the MultiView imaging mode.
•
In 3D programs the exposure values are given separately for
each image resolution. The image resolutions that are not
available are shown with faded buttons. Select also the ULD
(Ultra Low Dose) button if you wish to adjust the presets for the
ULD function.
5. Use the minus or plus buttons to set the exposure values you wish to
use.
6. Select the green check mark button.
7. Repeat for an other program type, patient size or image resolution
(3D) if needed.
8. Select the green check mark button.
NOTE
Always try to minimise the radiation dose to the patient.
NOTE
You can restore the exposure values that have been preset at the factory
(i.e. overrule your own settings) by selecting Program > 2500 Reset to
Factory Defaults.
NOTE
You can adjust the preset exposure values temporarily as described in
section "Adjusting exposure values for current exposure" on page 37.
11 Settings
78 Planmeca ProMax
User's manual