6.10.1 X-ray unit up / down
Use the X-ray unit up and down buttons to adjust the X-ray unit to suit the
height of the patient.
The X-ray unit moves slowly at first, then faster.
NOTE
If for some reason either of the buttons gets stuck during operation, you can
stop the up / down movement by pressing any of the other control buttons or
the positioning joystick. This is a safety measure that guarantees that the
up / down movement can be stopped in an emergency.
NOTE
Be careful that the X-ray unit does not hit the ceiling when you press the up
button. The maximum height can be adjusted to suit offices with a low
ceiling. Contact your service technician for help.
NOTE
Make sure that there is no object under the X-ray unit when you press the
down button. If something is in danger of becoming trapped, release the
button immediately to stop the movement. Clear any obstruction before
pressing the button again.
NOTE
The column movement stops automatically if the emergency stop plate at
the bottom is pressed upwards. Clear any obstruction before moving the
column again.
NOTE
When positioning seated patients (for example, in a wheelchair) always first
move the X-ray unit down before you position the patient in the unit.
6 Main parts
20 Planmeca ProMax
User's manual