Turn this option ON if you wish to use the midsagittal and Frankfort
plane lights when you take 3D exposures.
•
Temple Supports
Turn this option OFF if you do not wish to use the temple supports.
•
Return C-arm Automatically
Turn this option ON if you wish to set the X-ray unit so that the C-arm
will automatically return to the starting position at the end of an
exposure cycle. Note, however, that the automatic function works
only if the exposure button is pressed and held down for the entire
duration of the exposure.
3. Select the green check mark button.
•
To control exposure settings:
1. Select User > 1300 Operational Settings > 1340 Exposure Settings.
2. Select:
•
Continuous Exposure Indicator
Turn this option ON if you wish to set the X-ray unit so that the yellow
radiation warning light stays on continuously (instead of flashing)
when you take an exposure.
Note that the setting affects all indicator lights that are connected to
your X-ray system (indicator light on the hand-held exposure switch,
wall exposure switch(es) and remote exposure lamp).
3. Select the green check mark button.
11 Settings
User's manual
Planmeca ProMax 75