Installing a CryoCyl Low Pressure Cylinder
Wear protective clothing, including cryogloves, goggles, boots and a
protective apron.
The liquid-nitrogen supply pressure must not exceed 2.0 bar (30 psi).
If the hose is not already fitted to the Kryo550-16, wrap the male hose thread with PTFE
tap and fix to the LN
2
input on the back of the chamber. Tighten securely with a spanner.
Connect an appropriate adapter to the cylinder, observing the manufacturer’s instructions,
then connect the delivery tube to the adapter ensuring that the fibre washer is fitted to this
joint. Tighten securely with a spanner.
After connecting both ends of the delivery tube, open the liquid-nitrogen tap
slowly and check for leaks.
2-2 Installing the Kryo550-16