EMC and ESD PRECAUTIONS
The following precautions must be taken to ensure that the equipment is not
damaged by electrostatic discharge (ESD), and that its immunity to radio
frequency interference is not compromised.
Ensure that the Kryo550-16 is not running a program and that the mains power is
disconnected before connecting or disconnecting any cable.
Fit the plastic covers and plugs provided to all unused connectors.
Immediately before touching the sample PRT, touch an earthed part, such as the metal
surface at the rear of the Kryo550-16.
Do not touch any uncovered connectors with your hands or tools, even when the
equipment is switched off.
When connecting the system to a PC via the 9 way plug, always use a fully screened cable
no longer than 2m.
Take care to avoid placing freezers in environments influenced by sources of
electromagnetic interference, such as cyclotrons, large transformers etc.